TM 11-5840-281-12-1
4-47.
PERFORMANCE TESTS. (CONT)
NOTE
If MONITOR meter and multimeter indications differ consistently, MONITOR
meter is misalined. Refer to higher category of maintenance.
CONTROL SETTINGS
TEST PROCEDURE
INDICATION
On local control monitor,
On local control monitor,
MONITOR meter and multi-
place LOCAUREMOTE switch
connect positive lead of
meter indicate + 35 + 1 vdc. If
in RT position. Place MONITOR
multimeter to VOLTAGE MON
indications are Incorrect,
switch to + 35V position. Set
J5303 and connect negative
up multimeter to measure
lead to GND J5304.
+35 v.
Place MONITOR switch to + 28V
Same as above.
MONITOR meter and multi-
position.
meter indicate + 28 + 2 vdc. If
indications are incorrect,
refer to higher category of
maintenance.
Place MONITOR switch to + 15V
Same as above.
MONITOR meter and multi-
posItion.
meter indicate + 15 + 1 vdc. If
indications are incorrect,
refer to higher category of
maintenance.
Place MONITOR switch to + 12V
Same as above.
MONITOR meter and multi-
posItion.
meter indicate + 12 + 1 vdc. If
indications are incorrect, see
Place MONITOR switch to - 35V
Connect positive lead of
MONITOR meter and multi-
position. Set up multimeter
meter indicate - 35 + 1 vdc. If
to measure - 35 vdc.
connect negative lead to
indications are incorrect, see
VOLTAGE MON J5303.
Place MONITOR switch to - 12V
Same as above.
Monitor meter and multimeter
position.
indicate - 12+ 1 vdc. If indi-
cations are incorrect, see
4-145