TM 11-6625-847-45
Item
No.
Indication
Probable trouble
Procedure
2
Amplitude of 1-usec
Defective AMPLITUDE con-
Replace defective potentiometer
pulse output does not
trol R3.
R3.
vary when potentiom-
eter R3 is rotated
from fully counter-
clockwise to fully
clockwise
3
Pulse width of 160-usec
Defective pulse WIDTH con-
Replace defective potentiometer
and 85-usec pulse
trol R2.
R2.
outputs does not vary
when potentiometer
R2 is rotated from
fully counterclock-
wise to fully clockwise.
4
Pulse width of 1-usec
Defective pulse WIDTH control
Replace defective potentiometer
pulse output does
R4.
R4.
not vary when poten-
tiometer R4 is rotated
from fully counter-
clockwise to fully
clockwise.
5
The AN/USK-140(0)
Elective output coupling ca-
Replace defective capacitor A1C7.
does not show 85-usec
pacitor A1C7.
pulse output at connec-
tor J3, but 150-usec
pulse output is ob-
servable at connector
J2.
6
The AN/USM-140(*)
Defective output coupling ca-
Replace defective capacitor A14.
does not show 160-
pacitor A1C6.
usec pulse output at
connector J2, but 85-
usec pulse output is
observable at con-
nector J3.
7
The 150-, 86-, and 1-
Defective pulse shaping cir-
a. Connect test equipment as
usec pulse outputs at
cuitry (transistors Q1 and Q2).
described in paragraph
connectors J2, J3,
and J4 are not visible
waveshapes.
on the AN/USM-
140(*).
b. Connect the AN/USM-140(*)
probe to base of transistor
Q1, and adjust the AN/
USM-140(*) to show a
500 cps audio signal. If
signal is not present, check
capacitor A1C1, resistor
for defects. Replace
defective parts.
c. If audio signal is present at
base of transistor Q1, con-
nect the AN/USM-140(*)
probe to collector of tran-
sistor Q1 and adjust the
AN/USM-140(*) to show
2-11